Abstract

We report the first successful application of corona charging noncontact C-V and I-V metrology to interface and dielectric characterization of high-k/III-V structures. The metrology, which has been commonly used in Si IC manufacturing, uses incremental corona charge dosing, ΔQC, on the dielectric surface, and the measurement of surface voltage response, ΔVS, using a Kelvin-probe. Its application to In0.53Ga0.47As with a high-k stack required modifications related to the effects of dielectric trap induced voltage transients. The developed Corona Charge-Kelvin Probe Metrology adopted strictly differential measurements using ΔQC and ΔV, and corresponding differential capacitance rather than measurements based on total global charge, Q, and voltage, V, values.

Electrical characterization data including interface trap density, electrical oxide thickness, and dielectric leakage are presented for a sample containing an In0.53 Ga0.47 As channel overlaid with a bilayer (2nm Al2O3/5nm HfO2) dielectric stack that is considered to be very promising for application in performance NFETs with high-mobility channels.

Topic

corona charge, dielectric characterization, non-contact C-V, non-contact I-V